MouseOx® Murine Pulse Oximeter System from Starr Life Sciences Corp

The World's first and only Pulse Oximeter for mice and rats

Specifications:

Oxygen Saturation (Pulse rates 90 to 900 beats per minute, BPM)

Arterial Oxygen Saturation is reported to the user after each heartbeat. Every 0.72 sec, the screen is refreshed to display all the oxygen saturation values measured since the previous screen refresh.

Pulse Rate

Rate (BPM)

Resolution (BPM)

210

2.4

300

4.9

400

8.7

500

13.5

600

19.4

700

26.2

800

34.0

900

42.9

Breathe Rate

Rate (BrPM)

Resolution (BrPM)

25

<0.1

100

0.6

200

2.2

300

4.9

400

8.7

Pulse Distention

A measurement of the distention of the arterial blood vessels between the sensor pads. Provides a non-invasive indication of blood flow.

Breathe Distention

Measurement Response Time: Breath distention is reported to the user every 1.7 seconds.  The value reported is the result of a moving average of the rate of breathing for the 10 previous breath rate measurements.  Note that the screen still updates at a rate used for the other parameters (0.72 sec), but new breath distention data are presented at a lower rate.

Computer Requirements:
Processor:       
PC with Pentium™-Class processor (Pentium 1GHz or higher)
CD-ROM Drive
VGA or higher resolution monitor (Super VGA recommended)
 2.0 USB Port
Operating system:   Windows 2000, XP, and Vista
Memory:     512 MB RAM
5 MB Hard-Disk Space (Does not include data files)

Electrical Requirements:

Operating Wall voltage: 100-240 VAC @ 50-60 Hz
Device Operating Voltage: 12VAC
Max Operating Current: 500 mA

 